Continuous-Wave Doppler (CWD) Piggy Assembly

6497597
6497597-2
New
Ultrasound
Other Ultrasound
GE HealthCare
GE HealthCare
GE HealthCare
Outright
Add to Wish List
Sign In & Add To Cart
Sign In & Add To Cart
Please select date
Second Opinion Approval Needed

Enter your approval number and submit to add item(s) to cart.

Approval Number

Please enter approval number

OR

Don't know your approval number? Call 800-437-1171
Enter opt 1 for the first three prompts, and have your System ID available.

If you add item(s) to cart and submit your order without the
approval number, GE will contact you before your order
can be confirmed for shipment.

Second Opinion Approval Needed

Select your approver's name and submit to add item(s) to your cart

Approver Name

Please Select Approver Name

OR

Don't know your approval number? Call 800-437-1171
Enter opt 1 for the first three prompts, and have your System ID available.

If you add item(s) to cart and submit your order without
selecting an approver, GE will contact you before your order
can be confirmed for shipment.


Features

  • Compact Design
  • Field Replaceable Unit (FRU)
  • Good dimensional stability
  • Easy to install 20 Position, 2 Row, Push-Pull Connector

Product Overview

The Continous Wave Doppler Piggy Assembly is designed to enhance the Ultrasound Capabilities. The CWD board splits the transducer elements in equal halves where half of the elements are used as transmitters to produce a focused acoustic CWD transmit beam, and the other half as receivers to produce a focused receive beam. The signals applied to the transmit elements are square waves at the Doppler frequency of interest, typically 1.0MHz to 7.5MHz. The resultant "beamformed" CWD received signal is a combination of strong signals from stationary tissue, commonly referred to as clutter, and the much weaker Doppler signals from moving blood. This board uses high quality components which are assembled on a 6 layer FR-4 material PCB for reliable operation, efficient functioning and longer life span. The Piggy Assembly can be connected using the provided 20 Position, 2 row, push-pull type connector and can be installed over the PWB (Printed Wired Board) with the provided M3X4 philips cross recessed pan head machine screw and threaded stud. This CWD Piggy Assembly is compatible with GE's Voluson S10 Ultrasound systems. The item will be shipped in suitable packaging to provide adequate protection during transit.

  • Additional Features

    • 6 Layer, FR-4 PCB
    • CWD Technology
    • With Mounting Hardwares

    Equivalent Part :

    Below is more information on the equivalent parts. Parts without a hyperlink are listed for reference only and are not available for purchase online.

    Equivalent partPart Details
    6497597-2 CWD Piggy ASSY

    Return And Exchange

    GE HealthCare will accept, for return from customer, items that are in new condition, unworn, unaltered and free of damage. Items cannot be returned or exchanged beyond 30 days from customer’s receipt of the original order and all returned items will be subject to a 15% restocking fee. To obtain a full refund or exchange of the item within the 30 day return period (subject to the 15% restocking fee) customer must request a return materials authorization (RMA) from GE HealthCare and return the item to GE HealthCare utilizing this RMA number. Items must be returned in the original packaging, unused and free of damage.

    Please note, cuffs, BP items, sterile and hazmat items cannot be returned. If a replacement order is requested, the restocking fee may be waived at GE HealthCare’s sole discretion.